In vitro comparison of dental implant site preparation with ultrasonic system and conventional rotatory system

  • A. Lanata-Flores
  • , E. H.L. Nascimento
  • , D. R. Goulart
  • , F. Haiter-Neto
  • , M. de Moraes
  • , J. R. de Albergaria-Barbosa
  • Universidade Estadual de Campinas

Research output: Contribution to journal › Article › peer-review

1 Scopus citations

Abstract

Aim: To compare in vitro the site preparation of dental implant using a rotatory system and an ultrasonic system by evaluating the variability of temperature, time and primary stability that is provided by the different sequences of dental implant drills and tips. Materials and methods: Sixteen osteotomies were performed with different sequences of drills (Initial, Titamax 2.0, 2/3 and 3.0) for the rotatory group (RG) and tips (IM1S, IM2P, IM2/3, IM3P) for the ultrasonic group (UG) on bovine cortical-cancellous bone samples. Micro-CT images were acquired to assess bone microarchitecture and bone mineral density of each sample to guarantee similar bone conditions. For the RG, the osteotomies were performed with a dental implant contra-angle handpiece at 800 rpm and 50 N/cm maximum torque using an implant micromotor. For the UG, the osteotomies were performed using a Piezosurgery3 system with the respective tip. The temperature was measured and four implants 3.75 × 11 Titamax were installed in each group to measure the final torque. Results: For the RG, the following mean values were found: 22.87 s, 24.2°C superficial temperature, 27.6°C perforation temperature and 32.30 N/cm torque. The UG values were 72.9 s, 26.6°C superficial temperature, 32.1°C perforation temperature and 22.60 N/cm torque. Conclusions: Rotatory and ultrasonic systems are effective methods for implant site preparation with satisfactory outcomes. For cortical-cancellous preparation sites, the drills of the RG were related to lower temperatures and less time for osteotomy. Thus, the use of rotatory systems is preferred due to the heat differential generated during the dental implant sites preparation.

Original languageEnglish
Pages (from-to)108-116
Number of pages9
JournalOral Surgery
Volume13
Issue number2
DOIs
StatePublished - 1 May 2020

Keywords

  • dental implants
  • implant site preparation
  • piezosurgery
  • ultrasonic osteotomy

Fingerprint

Dive into the research topics of 'In vitro comparison of dental implant site preparation with ultrasonic system and conventional rotatory system'. Together they form a unique fingerprint.

Cite this